(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer states they have received replacement products, however, have not used them yet.Customer states she is in the hospital due to an infection and not her diabetes.
|
|
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on 3/19/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer states they have received replacement products, however, have not used them yet.Customer states she is in the hospital due to an infection and not her diabetes.
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 130-150mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on 03/02/2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 09/16/2018 and open vial date is 10/31/2017.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1 :445mg/dl, date: (b)(6) 2018, time:9:51am, fasting; result 2 :427mg/dl, date:(b)(6) 2018, time:7:34pm, fasting; result 3 :353mg/dl, date:(b)(6) 2018, time:6:17pm, fasting; result 4 :239mg/dl, date:(b)(6) 2018,time:1:00am,fasting; result 5 :537mg/dl, date:(b)(6) 2018, time:9:12pm, fasting.Customer called in states the meter is reading high.
|