TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer states they have received replacement products, however, have not used them yet.Customer states she is in the hospital due to an infection and not her diabetes.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 130-150 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 09/16/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (date/time not stored correctly): (b)(6).Customer called in states the meter is reading high.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on 3/19/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer states they have received replacement products, however, have not used them yet.Customer states she is in the hospital due to an infection and not her diabetes.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 130-150mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on 03/02/2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 09/16/2018 and open vial date is 10/31/2017.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1 :445mg/dl, date: (b)(6) 2018, time:9:51am, fasting; result 2 :427mg/dl, date:(b)(6) 2018, time:7:34pm, fasting; result 3 :353mg/dl, date:(b)(6) 2018, time:6:17pm, fasting; result 4 :239mg/dl, date:(b)(6) 2018,time:1:00am,fasting; result 5 :537mg/dl, date:(b)(6) 2018, time:9:12pm, fasting.Customer called in states the meter is reading high.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7373731
Report Number1000113657-2018-00363
Device Sequence Number13739721
Product Code NBW
UDI-Device Identifier00010939742445
UDI-Public(01)00010939742445
Combination Product (Y/N)N
Initial Reporter StateIN
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer03/02/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-