Model Number 8888340629
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the dressing was found to be wet and a crack was found at the device¿s connector.The catheter was replaced.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted that the device was received with a crack on the side of the devices connector.The condition in which the device was received precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the crack in the connector may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the dressing was found to be wet and a crack was found at the device¿s connector.It was reported that the device was tightened by hand.The catheter was replaced.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key7373835
Report Number3009211636-2018-00092
Device Sequence Number2312424
Product Code NIE
UDI-Device Identifier20884521057019
UDI-Public20884521057019
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K020089
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2021
Device Model Number8888340629
Device Catalogue Number8888340629
Device Lot Number1620300337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Initial Date Received by Manufacturer 03/09/2018
Supplement Date Received by Manufacturer08/09/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-