| Medical Device Problem Codes |
Moisture Damage (1405); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (case damage w/moisture) issue.The battery compartment was alleged to be cracked/damaged with moisture located inside the compartment.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission 11-apr-2018.Device evaluation: the device has been returned and evaluated by product analysis on 04-apr-2018 with the following findings: during investigation, the battery compartment was found to be cracked starting at the threads to the left side of the grip pad and from the case seal traveling to the grip pad.Moisture was observed in the battery compartment.A leak test was performed and a leak was identified in the battery compartment.The pump cover was opened and no further moisture was found.
|
| |
|