• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM08080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: as the device was not returned for evaluation, the reported device issue could not be reproduced.Based on lack of detailed information the nature of the reported device issue could not be determined.The investigation will be closed with inconclusive result.Labeling review: the current version of the instructions for use (ifu) supplied with this product was reviewed.Based on lack of detailed information the nature of the reported device issue could not be determined and it could not be verified if the potential risk or potential contributing factors are addressed in the document.However, the failure modes "perforation (.) detachment of part (.) stent graft fracture' were found addressed.In addition potential contributing factors for difficulties that may occur during deployment are addressed.The ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the stent allegedly fractured.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: as the device was not returned for evaluation, the reported device issue could not be reproduced.Based on lack of detailed information the nature of the reported device issue could not be determined.The investigation will be closed with inconclusive result.Labeling review: the current version of the instructions for use (ifu) supplied with this product was reviewed.Based on lack of detailed information the nature of the reported device issue could not be determined and it could not be verified if the potential risk or potential contributing factors are addressed in the document.However, the failure modes "perforation (.) detachment of part (.) stent graft fracture' were found addressed.In addition potential contributing factors for difficulties that may occur during deployment are addressed.The ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the material allegedly ruptured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
VASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7374396
Report Number2020394-2018-00284
Device Sequence Number13726295
Product Code PFV
UDI-Device Identifier04049519008509
UDI-Public(01)04049519008509
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/25/2019
Device Model NumberFEM08080
Device Catalogue NumberFEM08080
Device Lot NumberANAY3845
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/28/2018
Supplement Date Received by Manufacturer03/27/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-