• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR KNEE; EMPOWR 3D KNEETM FEMUR, NP, 9R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR KNEE; EMPOWR 3D KNEETM FEMUR, NP, 9R Back to Search Results
Model Number 241-02-109
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 03/20/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - after prolonged infection and several washouts with inserts swaps, the surgeon removed all primary implants and replaced the femur and tibial prosthesis with empower knee implants to serve as spacers until the infection subsided.When the infection cleared the surgeon removed the spacer implants and performed a successful expert knee revision.
 
Additional Manufacturer Narrative
Corrected data: corrected to document that this is the fourth revision.Manufacturer narrative: the reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 60 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Fourth revision surgery - after prolonged infection and several washouts with inserts swaps, the surgeon removed all primary implants and replaced the femur and tibial prosthesis with empowr knee implants to serve as spacers until the infection subsided.When the infection cleared the surgeon removed the spacer implants and performed a successful expert knee revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR KNEE
Common Device Name
EMPOWR 3D KNEETM FEMUR, NP, 9R
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7374667
Report Number1644408-2018-00267
Device Sequence Number1103680
Product Code JWH
UDI-Device Identifier00888912166768
UDI-Public(01)00888912166768
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/01/2023
Device Model Number241-02-109
Device Catalogue Number241-02-109
Device Lot Number953R1081
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/20/2018
Supplement Date Received by Manufacturer04/11/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date04/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
342-14-708, LOT 142T1011; 342-14-708, LOT 142T1011
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-