The reported unused device was returned for evaluation in its original unopened package.Label verification and visual inspection were performed.A hole was found in the poly material, adjacent to the device's distal obturator, of the package.This defect resulted in a sterile breach and was likely caused by the device after the sealing process.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of this product.The products released for distribution were found to have met all specifications prior to shipment.Of the lot containing (b)(4) units, no prior complaints have been received.A two-year review of complaint history revealed (b)(4) similar complaints, with (b)(4) devices, for this device family and failure mode.During this same time frame, (b)(4) devices have been manufactured and shipped worldwide, making the rate of occurrence of this (b)(4) percent.This reported packaging issue was obvious to the distributor, prompting the return of the device for evaluation.There was no patient involvement.As with all medical devices, examination of the product occurs multiple times prior to use.Good clinical practice would include examination and verification of the original packaging and its labeling to ensure both are intact.This issue will continue to be monitored through the complaint system to assure patient safety.
|