Model Number ZXT225
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Corneal Edema (1791); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt225 12.0 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2018.However, there is a planned explant scheduled for (b)(6) 2018 due to an unexpected post-operative refraction/myopia.Reportedly, the patient was very myopic to begin with, the iol is in perfect position, and the doctor repeated measurements multiple times.Through follow up explant of the lens was confirmed and there was no vitrectomy and no incision enlargement required, but one 10.0 nylon suture was placed.Reportedly, the replacement lens was the same model, but different diopter of 10.5.There was no patient injury, but the patient had significant corneal edema at their one day post-operative exam.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7375184
Report Number9614546-2018-00285
Device Sequence Number16410978
Product Code POE
UDI-Device Identifier05050474605633
UDI-Public(01)05050474605633(17)220629
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date06/29/2022
Device Model NumberZXT225
Device Catalogue NumberZXT225U120
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/26/2018
Supplement Date Received by Manufacturer06/06/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date06/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-