• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 3548214
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/28/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent primary breast reconstruction 550cc mentor cpx 4 breast tissue expander and experienced left sided deflation post procedure.As a result, the device was explanted on (b)(6) 2016.
 
Additional Manufacturer Narrative
The device history record (dhr) for the lot number 7333104 was reviewed on 5/18/2018 and it was verified that the device was manufactured in accordance with documented specifications and procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the returned product was completed by the failure analysis lab on 6/22/2018.Device evaluation summary: it was reported that a female patient underwent primary breast reconstruction 550cc mentor cpx 4 breast tissue expander and experienced left sided deflation post procedure.Upon receipt by mentor, the device returned without fluid.No foreign material was observed within the device.Brown material was observed on the shell surface.During examination, the product evaluation team observed a rent on patch.Microscopic examination of the edges of the rent gave no indications as to cause.No other anomalies were found.The complaint was confirmed since a rent was found on the device.The product evaluation team was unable to confirm any unusual failure mode, no corrective action will be taken at this time.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The device evaluation summary was erroneously submitted without analysis codes on 6/25/2018.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7375326
Report Number1645337-2018-01719
Device Sequence Number2311877
Product Code LCJ
UDI-Device Identifier00081317006558
UDI-Public00081317006558
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/04/2020
Device Catalogue Number3548214
Device Lot Number7333104
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2018
Supplement Date Received by Manufacturer02/27/2018
02/27/2018
02/27/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date06/06/2018
06/25/2018
06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-