• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CUFF STITCH 180 DEG LEFT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CUFF STITCH 180 DEG LEFT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 7209161
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the instrument broke inside the patient, all pieces were removed.No delay or patient injuries were reported.
 
Additional Manufacturer Narrative
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
 
Additional Manufacturer Narrative
One 180 degree cuff stitch left was returned for evaluation.Device is over four years old.Visual assessment confirmed the reported breakage.The distal tip at beginning of the first bend is broken off.The broken section was returned and shows evidence of plastic deformation at the break area.The condition of the device indicates it was subjected to excessive forces during use.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument's failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.No root cause related to the manufacturing process can be established.No further investigation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUFF STITCH 180 DEG LEFT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key7375412
Report Number1219602-2018-00377
Device Sequence Number11106048
Product Code NBH
UDI-Device Identifier03596010448927
UDI-Public(01)03596010448927(10)50523453
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign,health profe
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209161
Device Catalogue Number7209161
Device Lot Number50523453
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer04/16/2018
08/07/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date04/19/2018
08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
-
-