• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC THAL-QUICK ABSCESS DRAINAGE SET; GBX CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THAL-QUICK ABSCESS DRAINAGE SET; GBX CATHETER, IRRIGATION Back to Search Results
Catalog Number TQAS-1600-J
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
It was reported that the dilator of the thal-quick abscess drainage set became stuck in the drainage catheter.This device was removed, and the procedure was completed with another set.There were no injuries or additional procedures.The device was returned and examined, revealing that it was actually the dilator with the guidewire lodged inside of it, along with an unidentified hardened foreign matter.It was also found that the safety wire was broken, and the coils were elongated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THAL-QUICK ABSCESS DRAINAGE SET
Common Device Name
GBX CATHETER, IRRIGATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7377082
Report Number1820334-2018-00877
Device Sequence Number18864041
Product Code GBX
UDI-Device Identifier00827002063396
UDI-Public(01)00827002063396(17)181030(10)6311003
Combination Product (Y/N)N
Initial Reporter CountryGB
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTQAS-1600-J
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2018
Initial Report FDA Received Date03/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-