• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC HDM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC HDM; INTRAOCULAR LENS Back to Search Results
Lot Number 161307
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 02/27/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lens was returned in the vial enclosed in a biohazard bag.The cap and bung were present but no other lenstec packaging was returned.The inspection was performed at 10x magnification using a stereoscopic microscope.The lens was in two pieces, separated by the optic and when placed together, a complete lens was formed.The cross-sectional inspection revealed smooth and straight edges and appeared to be a cut made by a sharp instrument.This cut was probably made to facilitate removal of the lens from the eye.There were some scratches and an additional cut in the lens material around the area where the lens was separated.No other defects were noted.The dimension testing revealed that the lens was found to be within specification.A full audit of all batch documentation relating to the production of the lens was performed.The audit concluded that all procedures in manufacturing and packaging of the lens had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens were conducted correctly.In terms of the investigation, the defects seen were probably as a result of removing the lens from the eye.There was no evidence to suggest that any aspect of the lens was responsible for the surgical complication reported.Lenstec is therefore unable to determine a specific cause for this incident.Additionally, lenstec believes that the lens, its design and the manufacturing process are not at fault due to the extensive testing that we have performed on this model.Once any additional information is received a supplemental report will be submitted.
 
Event or Problem Description
Lenstec received via email a notification stating "patient was complaining after 2 months that they could not see very well.Doctor removed the lens and implanted another lens from another manufacturer.There was no adverse event with the patient.Patient complained of glare, rim artifact.".
 
Additional Manufacturer Narrative
The lens was returned in the vial enclosed in a biohazard bag.The cap and bung were present but no other lenstec packaging was returned.The inspection was performed at 1ox magnification using a stereoscopic microscope.The lens was in two pieces, separated by the optic and when placed together, a complete lens was formed.The cross-sectional inspection revealed smooth and straight edges and appeared to be a cut made by a sharp instrument.This cut was probably made to facilitate removal of the lens from the eye.There were some scratches and an additional cut in the lens material around the area where the lens was separated.No other defects were noted.The dimension testing revealed that the lens was found to be within specification.A full audit of all batch documentation relating to the production of the lens was performed.The audit concluded that all procedures in manufacturing and packaging of the lens had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens were conducted correctly.In terms of the investigation, the defects seen were probably as a result of removing the lens from the eye.There was no evidence to suggest that any aspect of the lens was responsible for the surgical complication reported.Lenstec is therefore unable to determine a specific cause for this incident.Additionally, lenstec believes that the lens, its design and the manufacturing process are not at fault due to the extensive testing that we have performed on this model.Medical monitor assessment has been added as supplemental information.
 
Event or Problem Description
Lenstec received via email a notification stating "patient was complaining after 2 months that they could not see very well.Doctor removed the lens and implanted another lens from another manufacturer.There was no adverse event with the patient.Patient complained of glare, rim artifact.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HDM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church, BB170 92
BB  BB17092
MDR Report Key7377244
Report Number9613160-2018-00017
Device Sequence Number11133395
Product Code HQL
UDI-Device Identifier00844369039338
UDI-Public00844369039338
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/17/2021
Device Lot Number161307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/27/2018
Supplement Date Received by Manufacturer02/27/2018
Initial Report FDA Received Date03/28/2018
Supplement Report FDA Received Date10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-