• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SINGLE LUMEN EMBOLECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SINGLE LUMEN EMBOLECTOMY CATHETER Back to Search Results
Catalog Number 1601-48
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
We have not received the device for evaluation since the device is still in the possession of the hospital.The surgeon confirmed that the device was used in a vessel with calcified plaque.Our ifu clearly/appropriately warns the users about the risks that could occur with the use of the embolectomy catheter including the risk of balloon rupture due to exposure to calcified plaque and over-inflated balloon.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent or calcified material ( eg.Chronic clot, atherosclerotic plaque ).This catheter is not designed to withstand the additional pull force needed to remove these materials.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.All of the (b)(4) units manufactured under this lot has been sold.Further, we have not received any other complaints of a similar a nature for devices from these lots.(b)(4).The current rate of occurrence is in line with our risk documents.There was no injury to the patient as the result of this incident.All of the fragments were successfully removed from the patient's artery by using an another catheter.
 
Event or Problem Description
Balloon ruptured during an embolectomy procedure in an iliac artery.Some balloon fragments remained in the iliac artery.So, the physician had to use another embolectomy catheter to remove the fragments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE LUMEN EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
MDR Report Key7377705
Report Number1220948-2018-00012
Device Sequence Number2312509
Product Code DXE
UDI-Device Identifier00840663100484
UDI-Public00840663100484
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/28/2023
Device Catalogue Number1601-48
Device Lot NumberSLC4317
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-