| Catalog Number |
KD692 SMN 10445205 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics has confirmed that dimension vista® ctni sample diluent (ctni sdil) lot 7bda34 may have an incomplete slit on the septum in the cap of the vial.A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the dimension vista system, due to a reduced volume of diluent being pipetted from the vial.Ctni sample diluent (ctni sdil) is used for dilutions of samples exceeding the analytical measurement range with the dimension vista® cardiac troponin i (ctni) flex reagent cartridge.Urgent medical device recall, vc-18-04.A.Us dated 21-mar-2018 and urgent field safety notice vc-18-04.A.Ous dated march 2018 were issued to customers who had been shipped dimension vista® ctni sample diluent (ctni sdil) lot 7bda34 to inform them of the issue.
|
| |
|
Event or Problem Description
|
|
A discordant elevated cardiac troponin i (ctni) result was obtained on a patient sample on the dimension vista 1500 instrument.The result was reported to the physician.The original sample was repeated and a lower result was obtained.A corrected result was reported.There are no known reports of patient intervention or adverse health consequences due to the discordant elevated ctni result.
|
| |
|
Search Alerts/Recalls
|