| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Most likely underlying root cause: mlc-20-user's test strip had poor storage (kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained of 151, 130 151 and 161mg/dl.The expected fasting blood glucose test result range is 90 to 120mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 11/30/2018 and open vial date is about 1 week ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Customer called about meter reading high blood results- there was a wide variance between the high results and the then next results.Customer claims the meter read 210mg/dl and then 120mg/dl on the next blood results.I could not verify those results in the memory.Customer did not verify if the results of 210mg/dl and 120mg/dl were fasting or non-fasting.
|
| |
|