Model Number GTS100R
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Date of Event 03/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The stent was not implanted in the proper location, but remains in the patient's eye therefore the surgery date was indicated in the implant date field.The stent remains in the patient's eye and it is not available for evaluation; therefore product testing on this device could not be performed.The device history records were reviewed for this manufacturing lot and there were no non-conformities found to be related to the reported event.As part of the device history records review the sterilization records for this lot were reviewed and this device lot passed sterility test.In addition, a complaint history review was completed.There were no complaints identified related to the reportable event.A review of the device labeling was completed.Malpositioned stent is identified in the device labeling as a known inherent risk of trabecular micro-bypass stent surgery.The ifu adequately provides instructions, precautions, and warnings for the proper use and handling of the device.Based on the results of the investigation, quality product deficiency was not identified.The omitted information was requested but was not provided.(b)(4).
 
Event or Problem Description
The surgeon reported that he was unable to implant a trabecular micro-bypass stent.The stent was lost in the eye during the surgical procedure and was not removed from the patient's eye.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: describe event or problem.The omitted information was requested but was not provided.(b)(4).
 
Event or Problem Description
Clinical follow-up was requested and on 04/16/2018 the surgeon provided the following additional details; the istent procedure of the right eye was aborted due to patient movement.The patient status and prognosis was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT TRABECULAR MICRO-BYPASS
Common Device Name
INTRAOCULAR PRESSURE LOWERING IMPLANT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
MDR Report Key7378668
Report Number2032546-2018-00029
Device Sequence Number1446231
Product Code OGO
UDI-Device Identifier00853704002005
UDI-Public00853704002005
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P080030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberGTS100R
Device Catalogue NumberGTS100R
Device Lot Number112966
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer04/16/2018
Initial Report FDA Received Date03/28/2018
Supplement Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
-
-