MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3235-40Q
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem Failure of Implant (1924)
Event Date 03/10/2015
Event Type  Injury  
Manufacturer Narrative
Correction: manufacturer report 2938836-2015-04224, should not have been reported as a medical device report (mdr) as the product has not been approved by fda and is part of investigation device exemption (ide).
 
Event Description
It was reported that the patient presented in-clinic for routine follow up.The device could not be interrogated.A surface ecg indicated that the device was not pacing.Patient noted having an mri in (b)(6) 2015.The device was explanted and replaced with a new competitive device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7379114
MDR Text Key103774118
Report Number2017865-2018-04501
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date05/31/2012
Device Model NumberCD3235-40Q
Device Catalogue NumberCD3235-40Q
Device Lot Number3267598
Other Device ID Number05414734503563
Date Manufacturer Received03/01/2018
Date Device Manufactured11/16/2010
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N