Manufacturer Narrative
|
Device was used for treatment, not diagnosis.
Please note that the reporter's name, phone number, and email address were not provided.
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.
The device was evaluated and it was observed that the device seized, jammed, and was generally moving heavily.
It was noted that the trigger hooked and ran heavily.
It was further determined that the device failed pretest for check the triggers and electric control unit (ecu) function.
Therefore, the reported condition was confirmed.
The assignable root cause was determined to be due to improper cleaning and maintenance, which is user error/misuse/abuse.
If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that small battery drive device was working inconsistently when the trigger was pressed.
It was not reported if the device was used in surgery, or if there was patient involvement.
It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.
It was not reported if there were any injuries, medical intervention or prolonged hospitalization.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|