• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.7MM CANNULATED DRILL BIT/QC 100MM; BIT,DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.7MM CANNULATED DRILL BIT/QC 100MM; BIT,DRILL Back to Search Results
Catalog Number 310.215
Medical Device Problem Codes Break (1069); Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that when using 2.4mm cannulated screw set and drilling over a 0.8mm guide wire and taking drill bit out after drilling, it was noticed that the tip of the drill bit had broken off.Fragments of the drill bit were left in patient.Also during this foot procedure, one of the 0.8mm guide wires was bent during insertion of a screw.The bent guide wire was taken out and replaced with a new one.There was 5-7 minutes delay in the surgery.Procedure was completed successfully.Patient status reported as stable.Concomitant devices reported: 2.4mm cannulated screw (part 211.820, lot number unknown, quantity 1) this report is for one (1) 1.7mm cannulated drill bit this is report 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.7MM CANNULATED DRILL BIT/QC 100MM
Common Device Name
BIT,DRILL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7379842
Report Number2939274-2018-51294
Device Sequence Number1657561
Product Code HTW
UDI-Device Identifier10886982186225
UDI-Public(01)10886982186225(10)LOTUNKNOWN
Combination Product (Y/N)N
Initial Reporter StateIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number310.215
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient Weight95
-
-