Model Number CRPLUS
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported issue.Physio determined that the cause of the reported issue was due to the on/off lid latch missing from the device.This prohibited the device from being able to power on.The customer received a replacement device.
 
Event or Problem Description
The customer contacted physio-control to report that the switch on their device was damaged and was keeping the device's charge-pak from going in and out.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the on/off lid latch was missing and the device could not be powered on.
 
Additional Manufacturer Narrative
Section d4 of the initial medwatch report indicates: device identifier (di) number ¿ blank.Section d4 of the initial medwatch report should indicate: device identifier (di) number ¿ (b)(4).
 
Event or Problem Description
The customer contacted physio-control to report that the switch on their device was damaged and was keeping the device's charge-pak from going in and out.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the on/off lid latch was missing and the device could not be powered on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR® PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7380522
Report Number0003015876-2018-00517
Device Sequence Number1051548
Product Code MKJ
UDI-Device Identifier00885074440323
UDI-Public00885074440323
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Initial Date Received by Manufacturer 03/01/2018
Supplement Date Received by Manufacturer01/16/2020
Initial Report FDA Received Date03/28/2018
Supplement Report FDA Received Date02/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-