MAUDE Adverse Event Report: ZIMMER BIOMET, INC. C/M ELBOW HUMERAL ASSEMBLY; PROSTHESIS, ELBOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. C/M ELBOW HUMERAL ASSEMBLY; PROSTHESIS, ELBOW Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Code Available (3191)
Event Date 01/19/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Report source, foreign - the event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Product location unknown.
 
Event Description
It was reported that the patient was revised due to a mechanical fracture of the humeral component.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to human factors; it was stated that "the patient was overweighting the operated extremity." if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C/M ELBOW HUMERAL ASSEMBLY
Type of Device
PROSTHESIS, ELBOW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7381335
MDR Text Key103823110
Report Number0001822565-2018-01858
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
PK001989
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/31/2012
Device Model NumberN/A
Device Catalogue Number32810502506
Device Lot Number60825785
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-