| Catalog Number |
HBD-W-8-9-10 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
| Date of Event |
03/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.The pre-loaded wire guide was included in the return.A visual inspection of the catheter showed a crack in the catheter approximately 150 cm from the distal end.No portion of the catheter appeared to be missing.The wire guide did not show any kinks or bends.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual product handling conditions could not be duplicated in the laboratory setting.This limits our ability to conclusively determine a cause.Damage to the catheter can occur if the device experiences excessive pressure during general handling.Prior to distribution, all hercules 3 stage wireguided balloons esophageal-pyloric-colonic are subjected to a visual examination to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
In preparation for a procedure, the user chose a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.The operator observed the outer sheath [catheter] was broken before advancing it into the endoscope.
|
| |
|
Search Alerts/Recalls
|