• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3889-33
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the lead model 3889-33 lot # va1mgrh showed part of the percutaneous extension was sticking out of the inner tray but was still inside the outer tray.Analysis found the inner lid was loose allowing component movement in the inner tray of the lead kit packaging.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a consumer regarding a patient who was implanted with a neurostimulator for unknown interstim therapy.It was reported that the lead¿s sterile packaging was opened in the or and the lead extension wire was inside the glued area.Sterility could not be guaranteed.Therefore, a new lead was opened and used for implantation.There were no further complications reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7381604
Report Number2649622-2018-02757
Device Sequence Number10795282
Product Code EZW
UDI-Device Identifier00643169887930
UDI-Public00643169887930
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/08/2021
Device Model Number3889-33
Device Catalogue Number3889-33
Device Lot NumberVA1MGRH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Initial Date Received by Manufacturer 03/28/2018
Supplement Date Received by Manufacturer07/11/2018
08/08/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date07/11/2018
10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-