MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. JELCO® VIAVALVE® SAFETY IV CATHETERS; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. JELCO® VIAVALVE® SAFETY IV CATHETERS; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number 326310
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
See mfr: 3012307300-2018-00833, 3012307300-2018-00879, 3012307300-2018-00880, 3012307300-2018-00881, 3012307300-2018-00882, 3012307300-2018-00883.
 
Event or Problem Description
It was reported that the smiths medical device "blood return big veins" and the device was hard to slide.It was indicated that a nurse blew a vein.No other information was provided.
 
Additional Manufacturer Narrative
One peripheral intravenous catheter was returned for analysis in used condition.Visual inspection of the catheter noted that the sub-assembly was not included so the valve could not be evaluated.The sample had some residue in the flash chamber and trace amounts in the noses as expected from normal use.The nose and cannula were measured to verify that they conformed to specification, as excessive clearance may result in blood leakage through the guard.The cannula and nose were within specification.Based on the evidence, the complaint was unable to be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO® VIAVALVE® SAFETY IV CATHETERS
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7381763
Report Number3012307300-2018-00883
Device Sequence Number13701366
Product Code FOZ
UDI-Device Identifier15019517078404
UDI-Public15019517078404
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K160235
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date07/08/2019
Device Catalogue Number326310
Device Lot Number3423954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer05/10/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date06/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-