| Catalog Number |
326310 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
See mfr: 3012307300-2018-00833, 3012307300-2018-00879, 3012307300-2018-00880, 3012307300-2018-00881, 3012307300-2018-00882, 3012307300-2018-00883.
|
| |
|
Event or Problem Description
|
|
It was reported that the smiths medical device "blood return big veins" and the device was hard to slide.It was indicated that a nurse blew a vein.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
One peripheral intravenous catheter was returned for analysis in used condition.Visual inspection of the catheter noted that the sub-assembly was not included so the valve could not be evaluated.The sample appeared clean and free of any residue.The nose and cannula were measured to verify that they conformed to specification, as excessive clearance may result in blood leakage through the guard.The cannula and nose were within specification.Based on the evidence, the complaint was unable to be confirmed.
|
| |
|
Search Alerts/Recalls
|