X-SPINE SYSTEMS, INC. FORTEX PEDICLE SCREW SYSTEM; PEDICLE SCREW SPINAL SYSTEM Back to Search Results
Model Number X022-0240
Medical Device Problem Codes Fracture (1260); Material Fragmentation (1261)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant notified the company that the wing lock pedicle screwdriver's threads disintegrated during surgery.The driver was inspected by the company and the damage was confirmed by the complaint coordinator.Functionality testing could not be performed due to the extent of the damage.The driver shows signs of repeated use.The damage to the threads could have been caused by over-torquing the handle attached to the driver.A dhr review was performed and no manufacturing anomalies were identified and the device met the required specifications.
 
Event or Problem Description
The complainant notified the company that the wing lock pedicle screwdriver's thread disintegrated during surgery.There are no known patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTEX PEDICLE SCREW SYSTEM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
X-SPINE SYSTEMS, INC.
664 cruiser lane
belgrade MT 59714
MDR Report Key7381948
Report Number3005031160-2018-00004
Device Sequence Number11055903
Product Code HXX
UDI-Device IdentifierM697X02202401
UDI-PublicM697X02202401
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K120832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX022-0240
Device Lot Number026792
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2018
Initial Report FDA Received Date03/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-