Model Number M003497320SR0
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that the coil (subject device) prematurely detached inside the patient during the procedure.The coil was retrieved along with the microcatheter using a goose neck snare device.The procedure was completed successfully using a different coil.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device was returned within the within a competitor microcatheter (prowler14) and goose neck snare.Visual inspection of the device revealed that the main coil was noted detached from the delivery wire and severely stretched, entrapped with a goose neck snare.There was procedural fluid noted within the microcatheter hub as well.During functional testing, the snare was unable to be removed from the returned microcatheter and the microcatheter was cut in an attempt to remove the snare, but it was not possible.After flushing the microcatheter, a 0.0158¿ patency mandrel was able to be advanced into the hub and proximal end of microcatheter without any difficulty.Information available indicated that the coil was confirmed to be in good condition prior to use.The presence of procedural fluid in the microcatheter entrapped with the coil and goose neck snare is an indication that continuous flush was not maintained.The lack of continuous flush could have caused the reported and analyzed issues; however, information available indicated that continuous flush was maintained and it cannot be determined if this continuous flush was present prior to medical intervention.Based on this information, the cause of the reported and analyzed anomalies could not be definitively determined.
 
Event or Problem Description
It was reported that the coil (subject device) prematurely detached inside the patient during the procedure.The coil was retrieved along with the microcatheter using a goose neck snare device.The procedure was completed successfully using a different coil.There were no reported clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIX2 ULTRASOFT SR 3MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7382255
Report Number3008881809-2018-00138
Device Sequence Number13479374
Product Code HCG
UDI-Device Identifier04546540691699
UDI-Public(01)04546540691699(17)190331(10)19015180
Combination Product (Y/N)N
Initial Reporter CountryHU
PMA/510(K) Number
K050700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Model NumberM003497320SR0
Device Catalogue NumberM003497320SR0
Device Lot Number19015180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2018
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROWLER MICROCATHETER(CODMAN)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-