Model Number EM210RF
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Summary of the analysis: analysis result indicates that the pretest temperatures were not able to perform due to the device was received dis-assembled, in parts.The parts were missing and the lever assembly on the housing was disassembled.Reassembled the lever assembly.The device was repaired, cleaned, tested and passed all manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Customer reported that during testing the unit was overheating.There was no patient injury.
 
Additional Manufacturer Narrative
Correction: serial number (b)(4).Analysis result indicates that customer¿s issue with overheating was not able to confirm.Pre-temperature tests were performed, the ambient temp or start temp was 75.6 and the raised or max temp was 89.0.The (max temp - the ambient temp = 13.4 which is passing.The max rise temp is 27 f for this drill.There was no fault found with the device.The device was repaired cleaned, tested, and passed manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE - STYLUS TOUCH®
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7382494
Report Number1045254-2018-00138
Device Sequence Number128211
Product Code HBE
UDI-Device Identifier00613994790217
UDI-Public00613994790217
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K081475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM210RF
Device Catalogue NumberEM210RF
Device Lot Number212461075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/06/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-