| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069); High impedance (1291)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of patient ecog suggests a lead break.Investigation of the explanted lead confirms a lead break, unable to determine root cause.Initial implant included the neurostimulator and four leads including three cortical strip leads and one depth lead.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017 during a review of neurostimulator data, an abnormal signal was observed on the depth lead connected to port 2 of the rns neurostimulator.Responsive stimulation was disabled on that lead and a decision was made to hold off on replacing the lead at that time.On (b)(6) 2018 the patient underwent a lead revision.
|
| |
|