| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lens was returned stuck in sfc-25 cartridge.Cartridge was returned in device tray with clear residue.Visual inspection found no visible damage to the device, residue on lens and the lens stuck in the cartridge.A lens work order search was performed.No other similar complaint type event found.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon loaded a 12.6mm micl12.6, implantable collamer lens, -11.5 diopter, and the lens ripped.There was no patient contact.The back up lens was used with no issues.The reporter stated that the cause of event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data concomitant medical products: cartridge model no.: sfc-25 and cartridge lot no.: 1380994.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: claim# (b)(4).
|
| |
|