MAUDE Adverse Event Report: CODMAN &SHURTLEFF, INC. CODMAN ESD 3 CSF EXTERNAL DRAINAGE SYSTEM WITH VENTRICULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN &SHURTLEFF, INC. CODMAN ESD 3 CSF EXTERNAL DRAINAGE SYSTEM WITH VENTRICULAR CATHETER Back to Search Results
Model Number 82-1730
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/27/2018
Type of Reportable Event Malfunction
Event or Problem Description
As reported on a spreadsheet received by natus on (b)(6) 2018, an eds3 (external drainage system) leaked at the proximal stopcock.Replaced with another; hospitalization of patient was extended.Additional information requested.
 
Additional Manufacturer Narrative
Additional information: patient information: no information received.No delay in surgery reported.This was not a revision of one or more implants.Three attempts were made for product followup.Natus has completed their investigation.The results forwarded by codman on (b)(6) 2018 (b)(4) included the referenced: methods: review of device history records.Review of complaints history.A review of complaint records for the previous 12 months could not be performed as no lot number information was provided.The most recent complaint and smt presentation was reviewed and did not show that this product family exceeded its upper control limit for complaints.Trends will be monitored for this or similar complaints.Investigation methods/evaluation results: complaint sample was not returned to codman and no lot number information has been provided; therefore, an evaluation of the device could not be performed and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.Root cause category: undetermined : insufficient information.Failure code codman : product analysis : device not returned root cause description: complaint sample was not returned to codman and no lot number information has been provided; therefore, an evaluation of the device could not be performed and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.No further corrective & preventative actions proposed.Document rationale for no corrective action: no corrective action required based on the above evaluation.Trends will be monitored for this or similar complaints.This complaint is now deemed closed.Should the product be returned, the complaint may be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN ESD 3 CSF EXTERNAL DRAINAGE SYSTEM WITH VENTRICULAR CATHETER
Common Device Name
EXTERNAL DRAINAGE SYSTEM
Manufacturer (Section D)
CODMAN &SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
MDR Report Key7383045
Report Number2023988-2018-00010
Device Sequence Number16448042
Product Code JXG
UDI-Device Identifier10886704040903
UDI-Public10886704040903
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K162437
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 03/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number82-1730
Device Catalogue Number821730
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer03/05/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date05/31/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-