ENCORE MEDICAL L.P. TURON SHOULDER; HEAD, HUMERAL, OFFSET, 50-18 Back to Search Results
Model Number 520-50-118
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a failed rotator cuff.The surgeon converted to a reverse.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a failed rotator cuff.The previous surgery and the revision detailed in this investigation occurred 1.1 year apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to this event.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to a failed rotator cuff.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
HEAD, HUMERAL, OFFSET, 50-18
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7383382
Report Number1644408-2018-00277
Device Sequence Number1052196
Product Code KWS
UDI-Device Identifier00888912025058
UDI-Public(01)00888912025058
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/23/2022
Device Model Number520-50-118
Device Catalogue Number520-50-118
Device Lot Number934C1124
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/21/2018
Supplement Date Received by Manufacturer04/19/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1684; 520-01-010, LOT 455G1237; 521-01-246, LOT 800N1067
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-