Model Number DBP-200SOLID145
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 02/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported oad was returned for analysis with the guide wire engaged in the device.Analysis of the handle revealed the saline sheath, driveshaft, and engaged guide wire to have been destructively cut.All cut sections exhibited adhered dried biological material.The cut and deformed driveshaft and guide wire sections are the result of the surgical intervention performed to remove the device from the patient.Further examination revealed the oad crown and guide wire spring tip to be intact and undamaged.When tested, the device functioned as intended with no abnormalities observed.There was no damage observed with the oad or guide wire sections that would have contributed to the reported event.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
During a peripheral atherectomy procedure using a csi orbital atherectomy device (oad), a dissection was noted.The target lesion was 60% stenosed and located in the distal superficial femoral artery (sfa).The lesion was treated using one pass with the oad at low speed and one pass at medium speed.Following the second pass with the oad, a dissection was noted and the oad became stuck in the patient.Surgery was performed to resolve the dissection and remove the oad, and the patient was stable following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key7383456
Report Number3004742232-2018-00084
Device Sequence Number13757048
Product Code MCW
UDI-Device Identifier10852528005275
UDI-Public(01)10852528005275(17)190930(10)203457
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133399
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/30/2019
Device Model NumberDBP-200SOLID145
Device Catalogue NumberDBP¿200SOLID145
Device Lot Number203457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight66
-
-