• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. SEEKER SUPPORT CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. SEEKER SUPPORT CATHETER Back to Search Results
Model Number SK9035M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during removal of the support catheter from the patient, the catheter allegedly detached into two segments.No further treatment was provided.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.One seeker catheter was returned in two segments for evaluations.The first segment measured 26.5cm in length and contained the distal tip of the device.The second segment was the remaining length of the catheter device.The break on the catheter was examined closer under magnification (20x) and the break was not clean with evidence of stretching and ridges.The condition of the break is indicative of excessive force used on the catheter.Therefore, the investigation can be confirmed for the alleged detachment of the device.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during removal of the support catheter from the patient, the catheter allegedly detached into two segments.No further treatment was provided.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER SUPPORT CATHETER
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7384010
Report Number2020394-2018-00303
Device Sequence Number1392904
Product Code DQY
UDI-Device Identifier00801741067952
UDI-Public(01)00801741067952
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK9035M
Device Catalogue NumberSK9035M
Device Lot NumberVTBU0507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer05/02/2018
Initial Report FDA Received Date03/29/2018
Supplement Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
-
-