| Model Number |
10185JP |
| Medical Device Problem Codes |
Fracture (1260); Stretched (1601); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/28/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This case was reviewed and investigated according to the manufacturer's policy.Additional information obtained regarding this therapeutic coronary procedure indicated resistance was met when trying to remove the device.The physician opted to use a balloon catheter to remove the wire to ensure the wire did not break off during removal.No tests/laboratory data was available.Other relevant history: no information was available.The implant or explant dates are not applicable to this device.A visual inspection was performed on the returned device.The wire connector was not returned.The distal coil was stretched to approximately 230 mm in length.The core wire at the distal coil was broken.Approximately 20 mm of the core wire extended past the sensor housing.The remaining broken piece of the core wire was attached to the dome.The broken core wire at the distal coil was observed to be twisted.All pieces of the device were present.A guide wire bend test could not properly be performed to duplicate the reported resistance and movement failure due to the damaged condition of the distal coil.As a result, the resistance and movement failures could not be confirmed by the investigation.The observed damage indicated that excessive stress was placed on the device.It is likely that the device became damaged during the removal process.The probable cause of the reported damage to the device is excessive stress placed on the device.Stress, impact, and forces associated with use can affect the integrity of the device.It is not possible to conclusively determine how the reported resistance or the damage to the device occurred.The instructions for use (ifu) warn to never advance, torque or retract a pressure guide wire which meets significant resistance.The ifu also cautions the pressure guide wire should not be advanced if resistance is encountered.The wire should never be forcibly pushed into a vessel.Any time that resistance is encountered, the wire should be withdrawn under fluoroscopic guidance.In some instances, the wire may kink and must be removed.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.Additionally, complaints are monitored by monthly complaint review board process.
|
| |
|
Event or Problem Description
|
|
It was reported the wire was recognized by the system, zeroed and normalized.Since the wire was difficult to control, the physician decided to change the wire.When removing the wire from the vessel, it came to a resistance.The wire was turned over and difficult to remove.The removal of the wire was performed using a balloon catheter.After extracting the wire, the doctor discovered that the tip of the wire was damaged.In order to control the situation, an x-ray image has been done and ensured that the wire was completely removed.The patient had no injuries and no consequences are to be expected.The procedure was completed with a new wire.No patient injury.Patient was discharged as expected; doing very well.Vessel: rca 11.This event is being reported because additional intervention was used to remove the device.
|
| |
|
Search Alerts/Recalls
|
|
|