| Model Number |
25199 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to the local facility but has not yet been received at the main office for evaluation.Once the device has been returned and investigated, a follow-up report will be submitted.(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported 1 led segment is broken on the display.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.When powered on, a segment on the red led display does not illuminate.The device was able to obtain readings and alarmed audibly and visually under alarm conditions.Internal inspection showed a defective component "d2" on the system circuit board causes the led segment to fail to illuminate.Initial reporter zip code exceeded maximum allowable digits, zip code is as follows: (b)(6).A service history record review reveals that this unit was in the field for over three (3) months with no previous reported issues related to this reported event.
|
| |
|