Model Number 010901
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during open nephrectomy procedure, while the device was being used to staple the vessels, the device was unable to fire and had poor staple formation.They also saw few pushing bars came up.They took a new loading to resolve the issue and complete the case.No patient injury.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.The visual inspection and functional evaluation of the device had acceptable results.A review of the device history record could not be performed because the lot number was not provided.However, records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE ENDO TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7385381
Report Number2647580-2018-01664
Device Sequence Number19180584
Product Code GDW
UDI-Device Identifier20884521047768
UDI-Public20884521047768
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K910192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number010901
Device Catalogue Number010901
Device Lot NumberP7G1398MX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date03/30/2018
Supplement Report FDA Received Date06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-