| Model Number |
H7493926020270 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr.: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent deployment difficulties occurred.The target lesion was located in the mid right coronary artery.A 2.75 x 20 synergy ii drug-eluting stent was deployed to treat the lesion.During deflation of the balloon, the first half of the artery was nice and open; however, the lower half looked like the stent had prolapsed or collapsed as there was kind of a narrowing.There was no further interventions performed and the procedure was completed.No patient complications were reported and the patient's status was fine.
|
| |
|