Model Number DFM8006SC
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation was inconclusive for a graft tear as no objective evidence has been provided to confirm any alleged deficiency with the device.It was unknown if procedural issues contributed to the reported event.Based on the available information, the definitive root cause is unknown.Labeling review: the current instruction for use (ifu) states: aggressive and/or excessive graft manipulation when tunneling, or placement within a too tight or too small tunnel, may lead to separation of the spiral beading and/or graft breakage.The distal anastomosis should be made after tunneling or suture disruption can occur.Do not pass the cuff portion (distal end) of the distaflo bypass graft through a tunneler sheath or the tissue tunnel, as this could lead to separation of the spiral beading and/or graft breakage.Distaflo bypass grafts do not stretch (are non-elastic) in the longitudinal direction.The correct graft length for each procedure must be determined by considering the patient's body weight, posture, and the range of motions across the anatomical area of graft implantation.Failure to cut the graft to an appropriate length may result in anastomotic or graft disruption, leading to excessive bleeding, and loss of limb or limb function, and/or death.When suturing, avoid excessive tension on the suture line, inappropriate suture spacing and bites, and gaps between the graft and host vessel.Failure to follow correct suturing techniques may result in suture hole elongation, suture pull-out, anastomotic bleeding and/or disruption.Size the graft appropriately to minimize excessive tension at the suture line.Use a tapered, non-cutting needle with a nonabsorbable monofilament suture approximately the same size as the needle.Take 2 mm suture bites in the graft following the curve of the needle and gently pull the suture at a 90° angle.Proper sizing of the graft length prior to implant will minimize suture hole elongation caused by excessive tension.To avoid extreme stress on the anastomosis and the graft, include the patient¿s weight and range of limb motion when determining graft length, tunnel length and location.To determine the correct graft length, drape the patient to allow full movement of the arm, shoulder girdle or legs.Cutting the graft slightly longer than necessary has been reported by some surgeons to further reduce the risk of stressing the graft or the anastomosis.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that while suturing the femoral (proximal) anastomosis site of the eptfe graft, the graft allegedly tore.It was further reported that the health care provider made revisions to the graft and completed the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRA EPTFE VASCULAR GRAFT
Common Device Name
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7386363
Report Number2020394-2018-00312
Device Sequence Number1252218
Product Code DSY
UDI-Device Identifier00801741000164
UDI-Public(01)00801741000164
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K983769
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDFM8006SC
Device Catalogue NumberDFM8006SC
Device Lot NumberVTAW0644
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2018
Initial Report FDA Received Date03/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight117
-
-