• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number L-12
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopy, when doing ligation, the thread broke.The nurse replaced the other lot of the device to complete the case.The patient outcome was unknown.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of three devices.Post market vigilance (pmv) led a photographic evaluation of device.The visual inspection of the returned photo noted that the suture was discolored, and appeared to have disintegrated within the retainer.The visual inspection of the returned product noted twelve pre-cut sutures and degraded sutures in a retainer.As the sealed products were expired, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturers¿ quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven 06473
MDR Report Key7386576
Report Number1219930-2018-01831
Device Sequence Number13820748
Product Code GAM
UDI-Device Identifier10884521045866
UDI-Public10884521045866
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2018
Device Model NumberL-12
Device Catalogue NumberL-12
Device Lot NumberA3C0841X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer05/03/2018
Initial Report FDA Received Date03/30/2018
Supplement Report FDA Received Date05/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-