| Model Number |
MMT-522NAP |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had compromised force sensor system alarm and drive support cap was protruded.The customer's blood glucose level was 115 mg/dl.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test.Unit was received with motor error alarm during the basic occlusion test due to loose/flush drive support disk.Unable to perform prime/compromised force sensor system test, excessive no delivery test and occlusion test or verify compromised force sensor system alarm due to motor error alarm.Insulin pump was received with loose drive support disk.
|
| |
|