AV-TEMECULA-CT HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM Back to Search Results
Catalog Number 1012068-07
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The armada 35 referenced is being filed under a separate medwatch report number.Evaluation summary: the device was returned for analysis.The reported difficult to position and difficult to remove were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the moderate to heavily tortuous, 100% stenosed anatomy as the guide wire was advanced resulted in the noted multiple sporadic core kinks and bends thus as the 4 x 150 mm armada 35 catheter was advanced resulted in the reported difficult to position and the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to the moderate to heavily tortuous, 100% stenosed, proximal to distal right superficial femoral artery (rsfa).After placement of a hi-torque (ht) versacore guide wire, a 4 x 150 mm armada 35 percutaneous transluminal angioplasty (pta) catheter was advanced, but met resistance with the guide wire and was only able to reach the proximal rsfa.The balloon of the pta catheter was inflated and deflated with no issues noted.During removal of the pta catheter from the anatomy, it again met resistance and became stuck on the guide wire.A non-abbott guide wire was placed to serve as a buddy wire so access was not lost and the ht versacore guide wire & armada pta catheter were removed from the anatomy as a unit.The procedure was completed with the successful stenting of the complete rsfa.There was no reported clinically significant delay in the procedure.There was no reported adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE VERSACORE MODIFIED J GUIDE WIRE SYSTEM
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7386990
Report Number2024168-2018-02364
Device Sequence Number10822494
Product Code DQX
UDI-Device Identifier08717648124211
UDI-Public08717648124211
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K083706
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Catalogue Number1012068-07
Device Lot Number7110961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2018
Initial Report FDA Received Date03/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight64
-
-