| Model Number |
FEM12080 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Perforation (2205); Misfire (2532)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer, however, one photo was provided and is currently under review.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the endovascular stent graft allegedly partially deployed during an a/v fistula graft thrombectomy venous anastomosis treatment.Therefore, the device was exchanged for another and the procedure was completed.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Investigation summary: on the basis of the returned stent graft delivery system and the provided photo, the alleged deployment failure was confirmed.The deployment mechanism was found in used condition, the stent graft was found loaded, and the outer sheath was found perforated by one stent graft strut, which made a stent graft deployment impossible.As a result of the investigation performed the complaint was confirmed.However, based on the available information and the evaluation of the returned sample, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risk.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend including the terminal cephalic arch or across the elbow joint has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." regarding the use of accessories the ifu states: "materials required for the fluency® plus endovascular stent graft procedure: introducer sheath with appropriate inner diameter".Furthermore, the ifu states: "do not kink the delivery catheter or use excessive force during delivery to the target lesion." and "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the endovascular stent graft allegedly partially deployed during an a/v fistula graft thrombectomy venous anastomosis treatment.Therefore, the device was exchanged for another and the procedure was completed.There was no reported patient injury.
|
| |
|