| Model Number |
0010-0622 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record confirmed that the device met all material, assembly and performance specifications prior to release.The returned fiber fiber did not exhibit signs of usage.Analysis revealed that the optical surface of the connector appeared burnt and melted.Information available indicated that no issues were encountered during testing of the fiber prior to use.Based on the information available and analysis of the device, the cause for the reported event cannot be determined.This report is for the same event as manufacturer report: 2937094-2018-00356.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was under anesthesia for a microlaryngoscopy procedure.Upon starting the procedure, the fiber connector end part (*subject device) connecting to the console, burnt.A second fiber was utilized to continue the procedure; however, the same issue occurred.The physician aborted the procedure after the second fiber failed.The patient was reported to be okay.This is the 1st of 2 reports filed for this event.
|
| |
|