• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES PRISM HBSAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES PRISM HBSAG Back to Search Results
Catalog Number 06D19-68
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.(b)(6).
 
Event or Problem Description
The customer reports that donor sample (b)(6) generated repeat (b)(6) results with the abbott prism hbsag assay lot 80049m500 when tested on abbott prism serial number (b)(4).This same sample had generated an initial (b)(6) result on abbott prism serial number (b)(4) with the same reagent lot.The customer does not perform confirmatory testing at their facility.The sample was therefore sent to another lab and tested with the abbott prism hbsag confirmatory assay and was confirmed (b)(6).The same sample tested (b)(6)with an nat procedure.No suspect results were reported from the lab.There is no impact to donor management reported.
 
Additional Manufacturer Narrative
Date rec¿d by mfr of the initial mdr documented the incorrect year.The correct year should be 2018 and not 2017.
 
Additional Manufacturer Narrative
The donors were contacted and asked to return for retesting.The results were non-reactive.Contamination and possible receipt of the hepatitis b vaccine were ruled out by the customer as possible reasons for the initial false reactive results.Master lot used in testing performed on (b)(6) 2018 was 82069m500.For reagent lot 80049m500, no other customer had registered a complaint.Therefore, no problematic complaint activity was identified that would indicate this product was performing contrary to claims.
 
Additional Manufacturer Narrative
A review of complaint tracking and trending metrics was performed and identified no related adverse or non-statistical trends in conjunction with the complaint issue currently under evaluation.No non-conformances, potential non-conformances or deviations were found, which may have impacted the performance of the product.A review of field data was performed.A total of 2,395,761 samples have been tested with prism hbsag reagent lot 80049m500 across different customer sites.This number of samples is greater than the minimum sample size of the assay package insert (25,238).The number of initial reactive results is 1,531 and the number of repeat reactive results is 322.The initial reactive rate: 0.06% (higher 95% confidence interval within package insert: 0.09%).The repeat reactive rate: 0.01% (higher 95% confidence interval within package insert: 0.06%).Assuming a zero prevalence of hbsag in the tested samples, the specificity is calculated to be 99.99%, which is within package insert specifications of 99.99% with a 95% confidence interval of 99.97% to 100.00%.Both the initial and repeat reactive rates as well as the specificity are within assay package insert specifications.The prism hbsag assay package insert provides information to address the current customer issue.Based on the available information from the customer site and the results of this evaluation, there is no evidence to reasonably suggest that a systemic issue or product deficiency exists.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISM HBSAG
Common Device Name
HBSAG
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
MDR Report Key7387622
Report Number1415939-2018-00069
Device Sequence Number11119425
Product Code KSJ
UDI-Device Identifier00380740062989
UDI-Public00380740062989
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Medical Technologist
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/25/2018
Device Catalogue Number06D19-68
Device Lot Number80049M500
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/12/2017
Supplement Date Received by Manufacturer04/03/2018
04/06/2018
04/30/2018
Initial Report FDA Received Date03/31/2018
Supplement Report FDA Received Date04/03/2018
04/09/2018
05/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABBOTT PRISM ANALYZER LN: 06A36-11; ABBOTT PRISM ANALYZER LN: 06A36-11; ABBOTT PRISM ANALYZER LN: 06A36-11; ABBOTT PRISM ANALYZER LN: 06A36-11; SN: (B)(4); SN: (B)(4); SN: (B)(4); SN: (B)(4)
-
-