• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number H965458890
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although it has been indicated that the device from the reported event will be returned to angiodynamics for evaluation, it has not yet arrived.Upon receipt of the sample and completion of the investigation, a supplemental medwatch will be submitted.(b)(4).
 
Event or Problem Description
As reported, a bioflo picc which had been inserted on (b)(6) 2017 was removed on (b)(6) 2018 due to fluid leaking when the catheter was flushed.Upon removal, a hole at the junction of the picc and the suture wing was noted.It has been indicated that the used catheter will be returned to angiodynamics for evaluation.
 
Additional Manufacturer Narrative
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode "catheter - hole distal to suture wing." no adverse trends were identified.The reported complaint description of a hole in the catheter is confirmed, as there is a hole in the catheter just distal to the suture wing.No manufacturing non-conformances or out of specification conditions were observed during sample evaluation.Although no definitive root cause could be determined from sample review, it is unlikely that the hole is a result of the device manufacturing as it was placed and in use for several months prior to the reported event.In addition, it is unlikely that this defect occurred at angiodynamics as minimal processing is performed on catheter prior to placing in tray for packaging.Manufacturing process controls for the bioflo picc devices include a 100% a visual inspection for molding defects and a guidewire insertion test to verify lumen patency.The guidewire must pass through the catheter with little to no resistance to be considered acceptable.In addition, all catheters are 100% sprint tested after the guidewire verification to ensure the catheter does not leak.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / BIOFLO
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7388017
Report Number1317056-2018-00035
Device Sequence Number196121
Product Code LJS
UDI-Device IdentifierH965458890
UDI-PublicH965458890
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K121089
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date08/31/2019
Device Catalogue NumberH965458890
Device Lot Number5238190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/15/2018
Supplement Date Received by Manufacturer03/15/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date07/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-