U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer's father reported via phone call that insulin pump alarmed for insulin pump error and button seemed to not be working.Customer¿s blood glucose was unknown at time of incident.Customer performed troubleshoot but did not resolve the issue.Customer advised to refer to back-up plan.Insulin pump will be return for analysis.
 
Additional Manufacturer Narrative
Pump received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime/delivery and excessive no delivery test or verify motion sensor test failure alarm due to blank display.Pump received with moisture damage motor, vibrator, keypad and battery tube assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7388077
Report Number3004209178-2018-61228
Device Sequence Number1252318
Product Code OYC
UDI-Device Identifier00643169479609
UDI-Public(01)00643169479609
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient Family Member or Friend
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAH
Device Catalogue NumberMMT-523LNAH
Device Lot NumberA3523LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age11 YR
Patient Weight42
-
-