| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device is undergoing an evaluation but has not yet been completed.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results the customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 125 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was performed non-fasting by the customer and produced test results of 177 and 175 mg/dl using true track meter.The test strip lot manufacturer's expiration date is 4/13/2020 and open vial date (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly): (b)(6).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose resultsthe customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 125mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was performed non-fasting by the customer and produced test results of 177 and 175mg/dl using true track meter.The test strip lot manufacturer's expiration date is 4/13/2020 and open vial date (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly).(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|