• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HI-TORQUE PILOT GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HI-TORQUE PILOT GUIDE WIRE Back to Search Results
Catalog Number 1010482-H
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the guide wire and guiding catheter were not coaxially aligned; thus interaction with the guiding catheter as the guide wire was advances resulted in the reported difficulty to position and the reported peeling of the coating.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the mildly tortuous, heavily calcified, concentric, 99% stenosed, distal right coronary artery (rca) two different non-abbott guide wires (gw) were used separately but failed to cross the lesion.A pilot gw was used but met resistance inside the guide catheter and was removed; outside the anatomy peeling of the coating was suspected.A different gw was used to complete the procedure without reported issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Subsequent to the initial medwatch report filed, additional information was received stating that there was no peeled coating.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7388657
Report Number2024168-2018-02371
Device Sequence Number1007727
Product Code DQX
UDI-Device Identifier08717648073892
UDI-Public08717648073892
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Catalogue Number1010482-H
Device Lot Number7091871
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer04/02/2018
Initial Report FDA Received Date04/02/2018
Supplement Report FDA Received Date04/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-