| Model Number |
Z9002 |
| Medical Device Problem Codes |
Bent (1059); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
02/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 and (b)(6) 2018.(b)(4).An attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a z9002 19.0 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2017.It was later explanted because a slight distortion or kink was noticed on the haptic and the lens would not center properly.The replacement lens was a back-up lens of the same model and diopter.Reportedly, the patient is doing fine post-operatively.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 04/03/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection showed loose fibers/particles on the lens related the handling of the unit out of a sterile environment.Residues of lubricant material were also observed on lens.Both haptics were observed damaged/distorted.One of the haptic was also observed detached.The detached haptic piece was not returned.The condition in which the sample returned is consistent with a product that was handled and prepared for a surgical process.The complaint issue reported was verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no additional investigation requests have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|