| Model Number |
8886660841 |
| Medical Device Problem Codes |
Component Missing (2306); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to use, when the box was opened, one needle came off and one suture didn't have a needle on it.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted one suture.The suture was observed to be detached from the needle; needle was not returned.The visual inspection and functional evaluation of the sealed samples had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturer's quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|